How to Perform an EKG and Spot Artifacts vs. Real Rhythms

대표이미지

📈 Is This Rhythm Actually Dangerous? [Part 2] — How to Perform an EKG and Tell Artifacts Apart

Illustration of three EKG artifact patterns — electrical interference, muscle tremor, and wandering baseline — compared to a real rhythm

Hi everyone :) Ever since that tooth-brushing artifact story from the last post, I've made it a habit to run to the patient first whenever I see a weird waveform. But it turns out there's a real knack to telling whether something's actually an artifact just by looking at the waveform. 60Hz interference, muscle tremor, and baseline wander — at first they all just looked like "messy waveforms" to me, but it turns out they each have a distinct look.


Today, let's cover how to actually perform an EKG, plus how to tell which artifact you're looking at when the waveform looks off.


I put this together by cross-referencing resources from Asan Medical Center, Seoul National University Hospital, the Korea Health Information Service, and LITFL.

 

 

📌 1. How to Perform an EKG (Electrode & Lead Placement)

No special prep is needed for the test, but smoking, drinking, caffeine, or vigorous exercise right before it can affect the results, so it's best to avoid them.

Step Details Practical Tip
① Prep the electrode sites Clean the site with an alcohol swab, and if there's a lot of body hair, shave the area before applying conductive electrode gel. Dead skin, oil, sweat, and body hair increase skin-electrode impedance, which produces a messy waveform contaminated with muscle-noise artifact.
② Attach the limb leads Attach to the right arm (RA), left arm (LA), left leg (LL), and right leg (RL, ground). Swapping left and right creates artifacts like lead I reversal, so make attaching them in the same order every time a habit.
③ Attach the chest leads Attach in the order V1 (4th intercostal space, right sternal border) → V2 (4th intercostal space, left sternal border) → V3 (midway between V2 and V4) → V4 (5th intercostal space, midclavicular line) → V5 (5th intercostal space, anterior axillary line) → V6 (5th intercostal space, midaxillary line). Getting the exact 4th/5th intercostal space right matters — being off by even one or two spaces can make it look like there's an ST-segment change that isn't really there, which can be mistaken for an MI.
④ Record Record for 3–5 minutes while the patient rests comfortably, without talking or moving. Talking or moving during the test easily introduces muscle-tremor artifact, so give the patient a heads-up beforehand.

Why prep the electrode site this way? Dead skin, oil, sweat, and body hair on the skin's surface all raise the impedance between the skin and the electrode. Higher impedance means the heart's tiny electrical signals don't come through cleanly, and muscle noise gets mixed in, making the waveform messy. Wiping off oil and dead skin with an alcohol swab and applying electrode gel lowers the impedance so the signal comes through clean. The reason the chest leads go in that specific order and position is similar — the precordial leads (V1–V6) are placed to view the heart from multiple angles around the horizontal plane, so the position has to be exact for each lead to actually capture the area it's meant to.


When I first learned to find the 2nd rib at the sternal angle (Angle of Louis) and count down precisely from there, I remember thinking, "does it really need to be this exact?" Later, once I learned that being off by even one intercostal space could make it look like there's an ST change that isn't actually there, I understood why we're told to count each intercostal space so carefully. Ever since, I've made it a habit to always start from the sternal angle and count down each time, rather than eyeballing it when placing the chest leads.

 

📌 2. EKG Artifacts — Types and How to Identify Them

Some distorted waveforms look just like a real arrhythmia but are actually caused by a technical issue. You can tell them apart to some extent just by how they look.

Type Characteristic Appearance Common Cause
60Hz (AC) interference A thick, uniform, regular tremor (60 cycles per second) overlaid on the baseline. Nearby electrical equipment like power cords or electronic devices, poor grounding
Somatic tremor (muscle tremor) Irregular, scattered, hair-fine spikes with varying amplitude and spacing. Patient movement, shivering from cold, being tense, Parkinson's disease
Wandering baseline The entire waveform slowly drifts up and down like a wave. Respiration, poor electrode contact, dried-out electrode gel
Lead reversal The waveform itself looks clean, but lead I is inverted and aVR is upright. Limb electrodes attached to the wrong side (left/right swapped)
Electrode pop A sudden vertical spike appears in just one lead. An electrode coming loose or momentarily detaching
CPR compression artifact A regular distortion that matches the rhythm of chest compressions. The mechanical force of compressions during CPR mixing into the signal
💡 Want to go deeper on this? I've put together nursing study guides that cover topics like this in more detail — check them out here.

💡 Lead I Reversal + Upright aVR — Worth Knowing

Of all of these, lead reversal is especially worth knowing about — if lead I looks flipped from normal and aVR is upright at the same time, that combination is a strong signal that it's not a real abnormal rhythm, but that the limb electrodes were likely attached backward.

→ Clinical action: If you don't recognize this combination and just chalk it up to "a weird rhythm," you can end up needlessly worried about a perfectly fine patient — recheck the electrode placement first, then reattach and re-record.

 

💡 If the Waveform Suddenly Looks Messy, Start by Replacing the Electrodes

Electrode gel tends to dry out over time once opened, so on the floor, the general rule of thumb is to replace electrodes every 48 hours for typical adults, and every 24 hours for higher-risk patients like older adults, neonates, or those with sensitive skin.

→ Clinical action: If the waveform suddenly gets messy, before jumping to suspect an arrhythmia, check how long the electrodes have been on — swapping in fresh ones is often a surprisingly quick fix.

 

 

📌 3. EKG Interpretation — What to Check Every Time

Here are four basics to check every time you run and read an EKG.

Item Details Practical Tip
Troubleshooting 60Hz noise Unplug nearby electronic devices or power cords if possible, ensure the device is plugged into a grounded wall outlet, and make sure the patient cable doesn't cross power cords. Turning on the device's built-in AC filter (50/60Hz filter) can also help clear up the baseline significantly.
Troubleshooting muscle tremor Provide a warm blanket if the patient is shivering from cold, help them relax their limbs comfortably, and place limb electrodes closer to the torso if tremor is severe. For patients with Parkinson's disease, having them tuck their hands under their buttocks can sometimes reduce resting tremor artifact.
Troubleshooting wandering baseline Instruct the patient to breathe quietly and regularly, wipe away sweat or skin oils thoroughly before attaching electrodes, and replace dried-out electrodes. Ensuring there is no excessive tension or pulling on the lead wires can also prevent the baseline from drifting with movement.

 

🎯 Wrapping Up

Here's today's takeaways summed up in one line:

Before jumping to assume an emergency just because an EKG waveform looks strange, get in the habit of first ruling out artifacts like 60Hz interference, muscle tremor, baseline wander, and lead reversal. In particular, the lead I reversal + upright aVR combination is a telltale sign of lead reversal — miss it, and you can end up needlessly on edge over a perfectly fine patient.


Next time you find yourself wondering, "Is this waveform actually a dangerous rhythm?" — I hope today's rundown gives you a solid starting point.


Nurses, you've got this today too 🙌

📚 Want more resources like this?

I've put together nursing study guides and quick-reference ebooks that pull all of this into one place — no more digging through fifteen tabs before your shift.

Browse the Study Guides →

 

📎 Disclaimer: This post is personal material compiled for educational purposes by cross-referencing resources from Asan Medical Center, Seoul National University Hospital, the Korea Health Information Service, and other Korean medical institutions. Actual clinical interpretation and treatment decisions must always follow your facility's protocol/policy and current clinical guidelines. Numbers and reference values reflect the publication dates of the cited sources and may vary by institution or guideline, so please separately confirm whether they've since been updated.

 

📌 References

· Electrocardiography (EKG) - Asan Medical Center Health Information

· ECG Lead Positioning - LITFL ECG Library Basics

· ECG Artifacts - LITFL ECG Library Basics

· Kligfield P, et al. Recommendations for the Standardization and Interpretation of the Electrocardiogram, Part I: The Electrocardiogram and Its Technology. Circulation. 2007

· Update to Practice Standards for Electrocardiographic Monitoring in Hospital Settings - Scientific Statement, American Heart Association. Circulation. 2017

Post a Comment

0 Comments