🩸 Thromboelastography (TEG/ROTEM) Deep Dive [Part 1] --- Principles and Parameters
Ever stood at the bedside of a massively bleeding patient whose PT, PTT, and platelet count all came back normal, while the wound site was still oozing steadily, and thought "the numbers say I'm fine, so why won't this stop bleeding"? And then, the first time you picked up a TEG or ROTEM report and stared blankly at unfamiliar abbreviations like R, K, angle, MA, and LY30, wondering "what is this test even measuring"?
PT, PTT, and INR only give you a single "time" --- how long it takes for the first fibrin strand to form in plasma. TEG and ROTEM, on the other hand, show the entire clotting process in whole blood as one continuous curve --- from clotting factor activation, through the final clot strength generated by platelets and fibrinogen, all the way to fibrinolysis, when that clot breaks back down. In today's Part 1, we'll build a solid foundation: the basic principles behind TEG and ROTEM, the normal ranges and biological meaning of each parameter, pitfalls in specimen collection and handling, and how these tests correlate with conventional coagulation studies.
Cross-referenced da Luz LT et al. Scand J Trauma Resusc Emerg Med 2013 (practical TEG application), Scarpelini S et al. Braz J Med Biol Res 2009 (TEG normal ranges), Lang T et al. Blood Coagul Fibrinolysis 2005 (ROTEM normal ranges), Whiting D, DiNardo JA. Am J Hematol 2014, and Wool GD, Carll T. Int J Lab Hematol 2023.
📌 Table of Contents (Click to jump immediately)
- 📌 1. The Basic Principle of TEG/ROTEM --- How Is It Different From PT/PTT
- 📌 2. TEG Parameters and Normal Ranges --- R, K, Angle, MA, LY30
- 📌 3. ROTEM Parameters and Channels --- EXTEM, INTEM, FIBTEM, APTEM, HEPTEM
- 📌 4. What Each Parameter Actually Reflects --- The Trap of the "80/20" Myth
- 📌 5. Pitfalls in Specimen Collection and Handling
- 📌 6. How Hematocrit and Anemia Affect Results
📌 1. The Basic Principle of TEG/ROTEM --- How Is It Different From PT/PTT
| Item | Details | Clinical Takeaway |
|---|---|---|
| Whole blood vs. plasma | Traditional coagulation tests like PT, PTT, and INR measure a single point in time --- how long it takes for the first fibrin strand to appear in plasma that's been centrifuged to remove all cellular components. TEG and ROTEM, in contrast, place whole blood --- red cells, platelets, and all --- between a cup and a pin, and record the physical strength of the developing clot as a real-time curve as clotting begins and progresses (Wool GD, Carll T. Int J Lab Hematol. 2023;45(5):643-658). | Remember that this lets you see, in one shot, the actual clot strength generated by platelets and fibrinogen interacting with each other, and later, the fibrinolysis process as that clot breaks down. |
| Normal PT/PTT doesn't mean you're in the clear | A review citing a small study of 31 combat trauma patients reported that 64% had abnormal viscoelastic testing values (the original study used ROTEM) on arrival to the ED, while conventional coagulation tests like PT/PTT were abnormal in only 10% of that same patient group (da Luz LT et al. Scand J Trauma Resusc Emerg Med. 2013;21:29, citing Doran CM et al. J Trauma. 2010;69(Suppl 1):S40-S48). | These numbers come from a single small combat-trauma cohort, so it's hard to nail this down as a universal rule that applies to every trauma patient --- but hold onto the general takeaway: a situation where the lab numbers look normal but the patient is clinically still bleeding is a real thing. |
Item | Details | Clinical Takeaway |
Whole blood vs. plasma | Traditional coagulation tests like PT, PTT, and INR measure a single point in time --- how long it takes for the first fibrin strand to appear in plasma that's been centrifuged to remove all cellular components. TEG and ROTEM, in contrast, place whole blood --- red cells, platelets, and all --- between a cup and a pin, and record the physical strength of the developing clot as a real-time curve as clotting begins and progresses (Wool GD, Carll T. Int J Lab Hematol. 2023;45(5):643-658). | Remember that this lets you see, in one shot, the actual clot strength generated by platelets and fibrinogen interacting with each other, and later, the fibrinolysis process as that clot breaks down. |
Normal PT/PTT doesn't mean you're in the clear | A review citing a small study of 31 combat trauma patients reported that 64% had abnormal viscoelastic testing values (the original study used ROTEM) on arrival to the ED, while conventional coagulation tests like PT/PTT were abnormal in only 10% of that same patient group (da Luz LT et al. Scand J Trauma Resusc Emerg Med. 2013;21:29, citing Doran CM et al. J Trauma. 2010;69(Suppl 1):S40-S48). | These numbers come from a single small combat-trauma cohort, so it's hard to nail this down as a universal rule that applies to every trauma patient --- but hold onto the general takeaway: a situation where the lab numbers look normal but the patient is clinically still bleeding is a real thing. |
I've had that moment myself, standing next to a massively hemorrhaging trauma patient with a lab slip showing normal PT, PTT, and platelets, thinking "then why isn't this stopping." In the moment, all I could do was note the mismatch between the numbers and what I was seeing and notify the provider right away --- it wasn't until we ran a TEG alongside it that I really understood, in a hands-on way, that a normal number doesn't mean normal coagulation.
📌 2. TEG Parameters and Normal Ranges --- R, K, Angle, MA, LY30
| Item | Details | Clinical Takeaway |
|---|---|---|
| R time, K time, alpha angle | R time (Reaction time) is how long it takes from when the sample is loaded until clotting is first detected, defined as reaching 2mm of amplitude; it reflects the enzymatic clotting factor cascade leading up to thrombin generation. K time (Kinetics) is how long it takes, after R time, to reach 20mm of amplitude, and the alpha angle expresses the slope of that same segment as a degree measurement --- both reflect how fast the initial clot is forming. In a normal-range study of 118 healthy adults, using citrated whole blood activated with kaolin, the reported ranges were R 3.8-9.8 min, K 0.7-3.4 min, and alpha angle 47.8-77.7° (Scarpelini S et al. Braz J Med Biol Res. 2009;42(12):1210-1217). | Interpret a prolonged R time as a problem in the clotting factor cascade, and a prolonged K time or reduced alpha angle as slower initial clot formation --- keep these separate. |
| MA, LY30 | MA (Maximum Amplitude) is the maximum amplitude the curve reaches, representing the final strength of the clot produced by platelets and fibrinogen. LY30 is the percentage decrease in amplitude measured 30 minutes after MA is reached, showing how much fibrinolysis has progressed. In the same study, MA ranged 49.7-72.7mm and LY30 ranged -2.3-5.77% (Scarpelini S et al. Braz J Med Biol Res. 2009;42(12):1210-1217). | These values vary somewhat across the literature depending on the lab, reagents, and patient population, so check your facility's own validated normal ranges first. |
| TEG 6s is a different generation from TEG 5000 | While the TEG 5000 converts mechanical resistance between a pin suspended on a torsion wire and a rotating cup into an electrical signal, the TEG 6s is a completely different generation of device that detects viscoelasticity optically, using an LED inside a cartridge to track the movement of the blood meniscus and measure resonance frequency (this is a resonance-frequency method, not ultrasound --- and it works on a fundamentally different principle from SEER sonorheometry, an ultrasound-based technology used in the Quantra device). A recent study validating a heparin-neutralization cartridge reported, for the kaolin-activated (CK) channel, MA 51.0-67.6mm and R 5.0-9.1 min, and for the heparinase-added (CKH) channel, MA 51.8-67.9mm, R 4.9-9.4 min, and LY30 0-3.2% (Hartmann J et al. Am J Clin Pathol. 2025;163(1):12-19). | Make sure in practice that you never mix and compare normal ranges between TEG 5000 and TEG 6s. |
Item | Details | Clinical Takeaway |
R time, K time, alpha angle | R time (Reaction time) is how long it takes from when the sample is loaded until clotting is first detected, defined as reaching 2mm of amplitude; it reflects the enzymatic clotting factor cascade leading up to thrombin generation. K time (Kinetics) is how long it takes, after R time, to reach 20mm of amplitude, and the alpha angle expresses the slope of that same segment as a degree measurement --- both reflect how fast the initial clot is forming. In a normal-range study of 118 healthy adults, using citrated whole blood activated with kaolin, the reported ranges were R 3.8-9.8 min, K 0.7-3.4 min, and alpha angle 47.8-77.7° (Scarpelini S et al. Braz J Med Biol Res. 2009;42(12):1210-1217). | Interpret a prolonged R time as a problem in the clotting factor cascade, and a prolonged K time or reduced alpha angle as slower initial clot formation --- keep these separate. |
MA, LY30 | MA (Maximum Amplitude) is the maximum amplitude the curve reaches, representing the final strength of the clot produced by platelets and fibrinogen. LY30 is the percentage decrease in amplitude measured 30 minutes after MA is reached, showing how much fibrinolysis has progressed. In the same study, MA ranged 49.7-72.7mm and LY30 ranged -2.3-5.77% (Scarpelini S et al. Braz J Med Biol Res. 2009;42(12):1210-1217). | These values vary somewhat across the literature depending on the lab, reagents, and patient population, so check your facility's own validated normal ranges first. |
TEG 6s is a different generation from TEG 5000 | While the TEG 5000 converts mechanical resistance between a pin suspended on a torsion wire and a rotating cup into an electrical signal, the TEG 6s is a completely different generation of device that detects viscoelasticity optically, using an LED inside a cartridge to track the movement of the blood meniscus and measure resonance frequency (this is a resonance-frequency method, not ultrasound --- and it works on a fundamentally different principle from SEER sonorheometry, an ultrasound-based technology used in the Quantra device). A recent study validating a heparin-neutralization cartridge reported, for the kaolin-activated (CK) channel, MA 51.0-67.6mm and R 5.0-9.1 min, and for the heparinase-added (CKH) channel, MA 51.8-67.9mm, R 4.9-9.4 min, and LY30 0-3.2% (Hartmann J et al. Am J Clin Pathol. 2025;163(1):12-19). | Make sure in practice that you never mix and compare normal ranges between TEG 5000 and TEG 6s. |
📌 3. ROTEM Parameters and Channels --- EXTEM, INTEM, FIBTEM, APTEM, HEPTEM
| Item | Details | Clinical Takeaway |
|---|---|---|
| How TEG and ROTEM parameter names correspond | TEG rotates the cup while the pin stays fixed; ROTEM works the opposite way, rotating the pin while the cup stays fixed --- which is why its parameters carry different names: CT (Clotting Time, corresponds to R time), CFT (Clot Formation Time, corresponds to K time), alpha angle, MCF (Maximum Clot Firmness, corresponds to MA), and ML (Maximum Lysis, corresponds to LY30). Only the names differ; the biological meaning each one reflects is identical to its TEG counterpart (Whiting D, DiNardo JA. Am J Hematol. 2014;89(2):228-232). | If you find yourself flipping back and forth between a TEG report and a ROTEM report, start by pulling up this name-correspondence table. |
| EXTEM, INTEM, HEPTEM, APTEM, FIBTEM | EXTEM is activated with tissue factor and reflects the extrinsic pathway, similar to PT; INTEM is activated with ellagic acid and phospholipid and reflects the intrinsic pathway, similar to APTT; HEPTEM adds heparinase to INTEM to neutralize any heparin present in the sample. APTEM is EXTEM with aprotinin (a fibrinolysis inhibitor) added, and FIBTEM is EXTEM with cytochalasin D added, which blocks cytoskeletal polymerization and pharmacologically disables platelet contractile function --- isolating fibrinogen's contribution on its own. | Remember that comparing INTEM and HEPTEM side by side lets you tell whether an abnormal finding is truly from a clotting factor deficiency or from heparin contamination/administration, and comparing EXTEM and APTEM lets you tell whether a weak clot is a structural problem or due to excessive fibrinolysis. |
Item | Details | Clinical Takeaway |
How TEG and ROTEM parameter names correspond | TEG rotates the cup while the pin stays fixed; ROTEM works the opposite way, rotating the pin while the cup stays fixed --- which is why its parameters carry different names: CT (Clotting Time, corresponds to R time), CFT (Clot Formation Time, corresponds to K time), alpha angle, MCF (Maximum Clot Firmness, corresponds to MA), and ML (Maximum Lysis, corresponds to LY30). Only the names differ; the biological meaning each one reflects is identical to its TEG counterpart (Whiting D, DiNardo JA. Am J Hematol. 2014;89(2):228-232). | If you find yourself flipping back and forth between a TEG report and a ROTEM report, start by pulling up this name-correspondence table. |
EXTEM, INTEM, HEPTEM, APTEM, FIBTEM | EXTEM is activated with tissue factor and reflects the extrinsic pathway, similar to PT; INTEM is activated with ellagic acid and phospholipid and reflects the intrinsic pathway, similar to APTT; HEPTEM adds heparinase to INTEM to neutralize any heparin present in the sample. APTEM is EXTEM with aprotinin (a fibrinolysis inhibitor) added, and FIBTEM is EXTEM with cytochalasin D added, which blocks cytoskeletal polymerization and pharmacologically disables platelet contractile function --- isolating fibrinogen's contribution on its own. | Remember that comparing INTEM and HEPTEM side by side lets you tell whether an abnormal finding is truly from a clotting factor deficiency or from heparin contamination/administration, and comparing EXTEM and APTEM lets you tell whether a weak clot is a structural problem or due to excessive fibrinolysis. |
📌 4. What Each Parameter Actually Reflects --- The Trap of the "80/20" Myth
| Item | Details | Clinical Takeaway |
|---|---|---|
| Is MA really 80% platelets, 20% fibrinogen? | It's commonly taught --- and even stated in StatPearls --- that MA's (MCF's) final strength is determined "80% by platelets, 20% by fibrin." But no solid original research has independently validated this exact ratio, and there's even a paper (Anesth Analg. 2015;121(4)) pointing out that the subtraction-based method used to separate platelet and fibrinogen contributions is itself methodologically flawed (Kaur J et al. Fibrinogen. StatPearls [Internet]). | Treat "80/20" as a rough educational estimate meant to give you a general sense, not an exact figure --- in reality, platelets and fibrinogen interact with each other to generate strength in a nonlinear way. |
| Is alpha angle purely a fibrinogen marker? | Alpha angle is frequently taught as a marker of fibrinogen function alone, but it actually reflects both fibrinogen and platelet function together --- so it can drop not only with hypofibrinogenemia, but also with thrombocytopenia or platelet dysfunction. | Don't jump to conclusions about the cause based on a single parameter --- build the habit of interpreting it together with the other parameters and channels. |
Item | Details | Clinical Takeaway |
Is MA really 80% platelets, 20% fibrinogen? | It's commonly taught --- and even stated in StatPearls --- that MA's (MCF's) final strength is determined "80% by platelets, 20% by fibrin." But no solid original research has independently validated this exact ratio, and there's even a paper (Anesth Analg. 2015;121(4)) pointing out that the subtraction-based method used to separate platelet and fibrinogen contributions is itself methodologically flawed (Kaur J et al. Fibrinogen. StatPearls [Internet]). | Treat "80/20" as a rough educational estimate meant to give you a general sense, not an exact figure --- in reality, platelets and fibrinogen interact with each other to generate strength in a nonlinear way. |
Is alpha angle purely a fibrinogen marker? | Alpha angle is frequently taught as a marker of fibrinogen function alone, but it actually reflects both fibrinogen and platelet function together --- so it can drop not only with hypofibrinogenemia, but also with thrombocytopenia or platelet dysfunction. | Don't jump to conclusions about the cause based on a single parameter --- build the habit of interpreting it together with the other parameters and channels. |
When I was first learning to read TEG reports, I used to jump straight to "the alpha angle is low, so fibrinogen must be low." Then I watched my preceptor check the platelet count right alongside it, and I remember thinking, "wait, why are you checking platelets for the alpha angle?" It was only later that I learned the alpha angle reflects both fibrinogen and platelet function together --- and ever since, I've made it a habit not to pin down a cause from just one parameter.
📌 5. Pitfalls in Specimen Collection and Handling
| Item | Details | Clinical Takeaway |
|---|---|---|
| Timing: non-anticoagulated vs. citrated samples | A whole blood sample drawn without anticoagulant needs to be tested within 4 minutes of the draw, whereas a citrated sample should NOT be tested immediately after the draw --- it needs about 30-40 minutes to stabilize before recalcification and activation in order to get reproducible results (da Luz LT et al. Scand J Trauma Resusc Emerg Med. 2013;21:29). | Testing remains valid for some period beyond the stabilization window, but that upper limit varies somewhat across the literature, so follow your facility's and equipment manufacturer's specific guidance first. |
| The 37°C heating trap --- hypothermic patients | TEG/ROTEM devices run the test after warming the sample to 37°C, regardless of the patient's actual body temperature. | Keep in mind that a hypothermic trauma patient's coagulation ability is actually further impaired by their low body temperature, but the report won't reflect that hypothermia effect --- so hypothermia itself needs to be corrected separately (we'll cover massive transfusion management more in Part 2). |
| Trace heparin contamination | The R value (TEG) or CT (ROTEM/INTEM) from non-anticoagulated whole blood can be sensitive enough to prolong even from trace heparin or from heparan sulfate that the body secretes naturally. | If the patient is on heparin, or the sample was drawn from a catheter, the rule of thumb is to run it side by side with a heparinase-containing channel (HEPTEM, or the TEG 6s heparin-neutralization cartridge) to distinguish a true clotting factor deficiency from a heparin effect. |
Item | Details | Clinical Takeaway |
Timing: non-anticoagulated vs. citrated samples | A whole blood sample drawn without anticoagulant needs to be tested within 4 minutes of the draw, whereas a citrated sample should NOT be tested immediately after the draw --- it needs about 30-40 minutes to stabilize before recalcification and activation in order to get reproducible results (da Luz LT et al. Scand J Trauma Resusc Emerg Med. 2013;21:29). | Testing remains valid for some period beyond the stabilization window, but that upper limit varies somewhat across the literature, so follow your facility's and equipment manufacturer's specific guidance first. |
The 37°C heating trap --- hypothermic patients | TEG/ROTEM devices run the test after warming the sample to 37°C, regardless of the patient's actual body temperature. | Keep in mind that a hypothermic trauma patient's coagulation ability is actually further impaired by their low body temperature, but the report won't reflect that hypothermia effect --- so hypothermia itself needs to be corrected separately (we'll cover massive transfusion management more in Part 2). |
Trace heparin contamination | The R value (TEG) or CT (ROTEM/INTEM) from non-anticoagulated whole blood can be sensitive enough to prolong even from trace heparin or from heparan sulfate that the body secretes naturally. | If the patient is on heparin, or the sample was drawn from a catheter, the rule of thumb is to run it side by side with a heparinase-containing channel (HEPTEM, or the TEG 6s heparin-neutralization cartridge) to distinguish a true clotting factor deficiency from a heparin effect. |
📌 6. How Hematocrit and Anemia Affect Results
| Item | Details | Clinical Takeaway |
|---|---|---|
| Hematocrit independently affects MA | A study that artificially reconstituted TEG samples while varying the proportions of blood components found that raising the hematocrit actually decreased clot strength as represented by MA, while adding more platelets increased clot strength --- confirming that hematocrit itself is a variable that independently affects TEG/ROTEM clot strength, separate from platelet count or function (Noorman F, Hess JR. Transfusion. 2018;58(10):2430-2436). | When interpreting results in a patient with severe anemia or a patient whose hematocrit has shifted rapidly from massive transfusion, factor in the effect of the blood composition itself. |
| Don't call MA/MCF alone a "platelet problem" | MA/MCF is a multifactorial parameter determined by a mix of variables --- fibrin formation, its interaction with platelets, Factor XIII-mediated fibrin stabilization, and hematocrit. | When a particular parameter comes back low, resist jumping straight to "platelet problem" --- interpret it together with other channels (like FIBTEM). |
Item | Details | Clinical Takeaway |
Hematocrit independently affects MA | A study that artificially reconstituted TEG samples while varying the proportions of blood components found that raising the hematocrit actually decreased clot strength as represented by MA, while adding more platelets increased clot strength --- confirming that hematocrit itself is a variable that independently affects TEG/ROTEM clot strength, separate from platelet count or function (Noorman F, Hess JR. Transfusion. 2018;58(10):2430-2436). | When interpreting results in a patient with severe anemia or a patient whose hematocrit has shifted rapidly from massive transfusion, factor in the effect of the blood composition itself. |
Don't call MA/MCF alone a "platelet problem" | MA/MCF is a multifactorial parameter determined by a mix of variables --- fibrin formation, its interaction with platelets, Factor XIII-mediated fibrin stabilization, and hematocrit. | When a particular parameter comes back low, resist jumping straight to "platelet problem" --- interpret it together with other channels (like FIBTEM). |
After watching a few patients on ongoing massive transfusion, I started noticing that every time fluids and blood products shifted the hematocrit up or down quickly, MA moved right along with it. The first couple of times, I just figured it was the platelet count fluctuating --- it wasn't until I saw the same pattern a third time that it really clicked that hematocrit itself independently affects MA. Since then, I've made it a habit to check hematocrit trends alongside TEG results in massive transfusion patients.
🎯 Wrapping Up
If I had to sum up today's post in one line:
TEG/ROTEM shows you clot strength and fibrinolysis in a way PT/PTT never can --- but you have to understand exactly what each parameter reflects and how to handle the specimen correctly before you can interpret the results properly.
In the next part, we'll take these concepts and apply them to real bedside scenarios --- trauma-induced coagulopathy, massive transfusion protocols, fibrinolysis phenotypes, and rebalanced hemostasis in cirrhosis. You've got this 🙌
📚 Want more resources like this?
I've put together nursing study guides and quick-reference ebooks that pull all of this into one place — no more digging through fifteen tabs before your shift.
Browse the Study Guides →📎 This post is intended to provide general medical information for clinical reference. Actual patient care and treatment must always follow the ordering provider's instructions and your facility's protocol/policy. Values and reference ranges reflect the publication dates of the cited references and may vary by laboratory, equipment, and guideline --- please verify against the most current updates separately.
📌 References
· da Luz LT, Nascimento B, Rizoli S. Thrombelastography (TEG®): practical considerations on its clinical use in trauma resuscitation. Scand J Trauma Resusc Emerg Med. 2013;21:29
· Doran CM, Woolley T, Midwinter MJ. Feasibility of using rotational thromboelastometry to assess coagulation status of combat casualties in a deployed setting. J Trauma. 2010;69(Suppl 1):S40-S48
· Scarpelini S, Rhind SG, Nascimento B, Tien H, Shek PN, Peng HT, Huang H, Pinto R, Speers V, Reis M, Rizoli SB. Normal range values for thromboelastography in healthy adult volunteers. Braz J Med Biol Res. 2009;42(12):1210-1217
· Hartmann J, Dias J, Shilo A, Bynagari Y, Garrett B, Jeske W, Manukyan Z, Mkhitaryan K, Adelmann D, Subramaniam K, Sakai T. TEG 6s coagulation testing with a novel heparin neutralization cartridge: Technical validation and determination of normal reference ranges. Am J Clin Pathol. 2025;163(1):12-19
· Lang T, Bauters A, Braun SL, Pötzsch B, von Pape KW, Kolde HJ, Lakner M. Multi-centre investigation on reference ranges for ROTEM thromboelastometry. Blood Coagul Fibrinolysis. 2005;16(4):301-310
· Whiting D, DiNardo JA. TEG and ROTEM: technology and clinical applications. Am J Hematol. 2014;89(2):228-232
· Wool GD, Carll T. Viscoelastic testing: Critical appraisal of new methodologies and current literature. Int J Lab Hematol. 2023;45(5):643-658
· Kaur J, Patel P, Jain A. Fibrinogen. StatPearls [Internet]. StatPearls Publishing
· Noorman F, Hess JR. The contribution of the individual blood elements to the variability of thromboelastographic measures. Transfusion. 2018;58(10):2430-2436; Solomon C, Ranucci M, Hochleitner G, Schöchl H, Schlimp CJ. Anesth Analg. 2015;121(4):868-878
0 Comments